Classification and cyclic fatigue evaluation of new kinematics for endodontic instruments

This study evaluated the effect of new motions of the motor TriAuto ZX2 on the cyclic fatigue of endodontic instruments. Vortex Blue 35.06 instruments were divided into four groups (n = 10) and tested for fatigue in a curved artificial canal (90° and 2 mm radius) using the following motions: continuous rotation (CR), Optimum Torque Reverse (OTR) set at 180° and the Optimum Glide Path (OGP), which was tested at 90° and 240°. The time to fracture (TTF) and the lengths of the fractured fragments were recorded.