Nellix endovascular aortic sealing endoprosthesis late explantation for concomitant type I endoleak and stent frames proximal caudal migration
Endovascular aneurysm sealing (EVAS) using the Nellix™ System was introduced in clinical practice with the aim of reducing the incidence of complications such as migration, endoleaks, and reinterventions after conventional endovascular aneurysm repair (EVAR). Although, initial efficacy data on this device have been encouraging, EVAS has also demonstrated to undergo adverse events. Herein, we report a case of Nellix graft explant due to endobags shrinkage after air bubble reabsorption leading to proximal type I A endoleak and stent migration.