Torque range, a new parameter to evaluate new and used instrument safety
The aim of the present study is to evaluate operative torque, torque at failure and the difference between these two values-the "torque range"-of two different NiTi files. We also sought to evaluate and compare these parameters for new and used files. Forty S-One 20.06 and forty M-Two 20.06 were tested, divided in four equal groups (n = 10) for each brand. Ten instruments from each brand performed three root canal treatments each on standardized extracted single-rooted teeth. Afterwards, each group was subjected to the following two tests: operative torque and torsional resistance. Mean values for all the tested groups were calculated. The results for new instruments showed that S-One developed significantly higher operative torque, but higher torsional resistance. The results for used instruments showed that the S-One 20.06 developed less operative torque and higher mean torsional resistance value compared to the M-Two. Moreover, the percentage reduction of both values was significantly higher for M-Two. The results of the present study showed a safer torsional behavior of the S-One. These results could be related to the heat treatment and the manufacturing process.