A comprehensive in vitro comparison of mechanical properties of two rotary endodontic instruments

01 Pubblicazione su rivista
Seracchiani M., Miccoli G., Reda R., Zanza A., Valenti Obino F., Bhandi S., Gambarini G., Testarelli L.
ISSN: 0976-6006

Aim: The aim of the present study was to compare two different nickel–titanium (NiTi) rotary files, F-One Blue and F2 Protaper Gold (PTG), evaluating their properties such as cyclic fatigue resistance, torsional resistance, and bending properties. Materials and methods: Sixty F-One Blue (25.06) and 60 PTGF2 have been randomly divided into 3 groups of 20 instruments each. Each group was tested in a different way, in order to analyze three different properties: cyclic fatigue resistance, torsional resistance, and flexibility. Cyclic fatigue resistance was evaluated by inserting the instruments for 16 mm in a stainless-steel artificial canal with a 90° angle and 5 mm of curvature with recommended speed and torque. Time to fracture was recorded using a chronometer (1/100 seconds). The number of cycles to fracture (NCF) was then calculated. Fragments were collected, and their lengths were measured. Torsional resistance was evaluated on the apical 3 mm of each of the files to calculate torque to fracture (TtF) using an endodontic motor (KaVo, Biberach, Germany), which automatically recorded the torque values every 1/10 s. All instruments were rotated at the same speed (300 rpm) and torque value (5.5 Ncm). Flexibility was evaluated using a calibrated load cell supported by a computer program. All the collected data were statistically analyzed (t-test) with a significance level set at 5%. Results: A significant difference in terms of cyclic fatigue resistance, torsional resistance, and flexibility between F-One Blue and F2 PTG was found (p < 0.05). The mean value of NCF was 604.16 (SD ± 23.32) for F-One Blue and 300.5 (SD 19.92) for F2 PTG. The mean value of TtF was 1.41 Ncm (SD ± 0.01) and 1.39 Ncm (SD ± 0.01) for F2 PTG. The mean value of bending test was 29 gcm (SD ± 1.15) for F-One Blue and 50 gcm (SD ± 2.30) for F2 PTG. Conclusion: F-One Blue better resists to flexural and torsional stresses and seemed to be more flexible. Since F-One Blue mechanical tested performances were better than the F2 PTG ones, these instruments should be considered a very promising instrument. Clinical significance: As evidenced by the results of this study, F-One Blue should be considered a very promising instrument that could improve endodontic clinical practice.

© Università degli Studi di Roma "La Sapienza" - Piazzale Aldo Moro 5, 00185 Roma