Contrast media injection protocol optimization for dual-energy coronary CT angiography. results from a circulation phantom

01 Pubblicazione su rivista
De Santis Domenico, Caruso Damiano, Schoepf U Joseph, Eid Marwen, Albrecht Moritz H, Duguay Taylor M, Varga-Szemes Akos, Laghi Andrea, De Cecco Carlo N
ISSN: 1432-1084

Objectives: To investigate the minimum iodine delivery rate (IDR) required to achieve diagnostic coronary attenuation (300 HU) with dual-energy coronary CTA. Methods: Acquisitions were performed on a circulation phantom with a third- generation dual-source CT scanner. Contrast media was injected for a fixed time whilst IDRs varied from 1.0 to 0.3 gI/s in 0.1-gI/s intervals. Noise-optimized virtual monoenergetic imaging (VMI+) reconstructions from 40 to 90 keV in 5 keV increments were generated. Contrast-to-noise ratio (CNR) and coronary HU were measured for each injection. Results: VMI+ from 40–70 keV reached diagnostic attenuation with at least one IDR. The minimum IDR achieving a diagnostic attenuation ranged from 0.4 gI/s at 40 keV (312.8 HU) to 1.0 gI/s at 70 keV (334.1 HU). Attenuation values reached with IDR of 1.0 gI/s were significantly higher at each keV level (p

© Università degli Studi di Roma "La Sapienza" - Piazzale Aldo Moro 5, 00185 Roma