On-chip real-time monitoring of multiple displacement amplification of DNA

01 Pubblicazione su rivista
Bruijns B. B., Costantini F., Lovecchio N., Tiggelaar R. M., Di Timoteo G., Nascetti A., de Cesare G., Gardeniers J. G. E., Caputo D.
ISSN: 0925-4005

The integration of systems for real-time monitoring of DNA amplification reactions is rather limited, because the heating system, the temperature control and the detection system are usually realized with separate modules. In this work, a lab-on-a-chip system for real-time monitoring of the multiple displacement amplification reaction (MDA) is presented. The amplification and detection unit consists of a system-on-glass (SoG) coupled to a microfluidic chip. The SoG includes thin film metallic resistors and amorphous silicon temperature sensors to control the temperature, as well as amorphous silicon photosensors and an interference filter for fluorescence detection on a single glass plate. The amplification reaction is carried out in the microfluidic chip made of cyclic olefin copolymer (COC). By using this setup, the multiple displacement amplification reaction can be monitored in real-time using the fluorophore [Ru(phen) 2 (dppz)] 2+ .